汪大海, 高远鹏, 罗欢. D8X射线衍射仪简易变温加热台的研制[J]. 分析测试技术与仪器, 2013, 19(2): 65-70.
引用本文: 汪大海, 高远鹏, 罗欢. D8X射线衍射仪简易变温加热台的研制[J]. 分析测试技术与仪器, 2013, 19(2): 65-70.
WANG Da-hai, GAO Yuan-peng, LUO Huan. Development of Simple Temperature Heating Station of D8 X-ray Diffractometer[J]. Analysis and Testing Technology and Instruments, 2013, 19(2): 65-70.
Citation: WANG Da-hai, GAO Yuan-peng, LUO Huan. Development of Simple Temperature Heating Station of D8 X-ray Diffractometer[J]. Analysis and Testing Technology and Instruments, 2013, 19(2): 65-70.

D8X射线衍射仪简易变温加热台的研制

Development of Simple Temperature Heating Station of D8 X-ray Diffractometer

  • 摘要: 成功研制的简易变温加热台,在不更换D8X射线衍射仪的附件情况下,能快速方便的安装在衍射仪上,进行X射线衍射原位变温测试,满足了从室温至600℃的原位变温X射线衍射测试需求.更换附件时间从2~3天缩短至30 min即可进行变温X射线衍射测试,提高了测试效率.同时,节省了购买同类变温加热台近20万元的资金.

     

    Abstract: A simples, variable temperature heating unit has been successfully developed,which be quickly and easily installed on the X-ray diffractometer,without any replacement of D8 X-ray diffractometer attachment cases. In situ variable temperature X-ray diffraction test,meeting the demand from room temperature to 600℃ in-situ temperature X-ray diffraction test can be carried out, shortening the time of replacement of the attachment from 2~3 days to 30 min. The test efficiency is improve. A fund of nearly 20 million has been saved from the purchase of a similar variable temperature heating device.

     

/