陈丽梅, 张慧, 白国涛, 马彩霞, 姚思雨, 王婧. 微波灰化-电感耦合等离子体发射光谱法测定婴幼儿乳粉中的钙和磷[J]. 分析测试技术与仪器, 2023, 29(4): 407-413. DOI: 10.16495/j.1006-3757.2023.04.010
引用本文: 陈丽梅, 张慧, 白国涛, 马彩霞, 姚思雨, 王婧. 微波灰化-电感耦合等离子体发射光谱法测定婴幼儿乳粉中的钙和磷[J]. 分析测试技术与仪器, 2023, 29(4): 407-413. DOI: 10.16495/j.1006-3757.2023.04.010
CHEN Limei, ZHANG Hui, BAI Guotao, MA Caixia, YAO Siyu, WANG Jing. Detemination of Calcium and Phosphorus in Infant Formula by Microwave Ashing- Inductive Coupled Plasma Emission Spectrometry[J]. Analysis and Testing Technology and Instruments, 2023, 29(4): 407-413. DOI: 10.16495/j.1006-3757.2023.04.010
Citation: CHEN Limei, ZHANG Hui, BAI Guotao, MA Caixia, YAO Siyu, WANG Jing. Detemination of Calcium and Phosphorus in Infant Formula by Microwave Ashing- Inductive Coupled Plasma Emission Spectrometry[J]. Analysis and Testing Technology and Instruments, 2023, 29(4): 407-413. DOI: 10.16495/j.1006-3757.2023.04.010

微波灰化-电感耦合等离子体发射光谱法测定婴幼儿乳粉中的钙和磷

Detemination of Calcium and Phosphorus in Infant Formula by Microwave Ashing- Inductive Coupled Plasma Emission Spectrometry

  • 摘要: 采用微波灰化-电感耦合等离子体发射光谱法(ICP-OES)测定婴幼儿乳粉中钙、磷元素含量. 采用微波灰化法对婴幼儿乳粉进行前处理,正交试验方法确定微波灰化最佳条件,灰化后产物用2 mL硝酸溶液(体积比为1∶1)溶解后,用ICP-OES对钙、磷元素进行含量检测. 磷加标回收率为86%~104%,钙加标回收率为87%~96%. 磷的相对标准偏差为2.5%~7.0%,钙的相对标准偏差为3.9%~10.0%,能够满足日常检测要求. 采用微波灰化法对婴幼儿乳粉中钙、磷元素进行样品前处理,相比微波消解方法,具有用时短、用酸量少、消解效果好、不需要进行赶酸处理等优点. 与干法灰化和湿法消解相比大大减少了样品处理时间. 采用微波灰化与ICP-OES结合对婴幼儿乳粉中的重要指标元素进行检测,在婴幼儿乳粉质量控制中有很好的应用价值.

     

    Abstract: The microwave ashing-inductive coupled plasma emission spectrometry (ICP-OES ) was used to determine the content of calcium and phosphorus in infant formula. The method of microwave ashing was used to pretreatment of the infant formula. The optimization conditions were determined by orthogonal test. After the microwave ashing, the ashes were dissolved with 2 mL nitric acid solution (volume ratio was 1∶1). The ICP-OES was used to determine the content of calcium and phosphorus. The recoveries of phosphorus and calcium were 86%~104% and 87%~96%, respectively. The relative standard deviation of phosphorus and calcium were 2.5%~7.0% and 3.9%~10.0%, respectively, which could meet the detection requirements. Compared with the method of microwave digestion, microwave ashing has the advantages of shorter time, less acid, better digestion effect and no need to drive acid treatment for calcium and phosphorus in infant formula. Compared with wet digestion and dry ashing, the sample processing time was greatly reduced. The detection of important index elements in infant formula by microwave ashing combined with ICP-OES has a good application value in the quality control of infant formula.

     

/