孙悦, 肖蓉. 岛津IR—408红外分光光谱仪故障二则[J]. 分析测试技术与仪器, 1996, (1): 50-51.
引用本文: 孙悦, 肖蓉. 岛津IR—408红外分光光谱仪故障二则[J]. 分析测试技术与仪器, 1996, (1): 50-51.
Sun Yue, Xiao Rong. Two Special Trouble Examples in Shimadzu IR-408 Infrared Spectrophotometer[J]. Analysis and Testing Technology and Instruments, 1996, (1): 50-51.
Citation: Sun Yue, Xiao Rong. Two Special Trouble Examples in Shimadzu IR-408 Infrared Spectrophotometer[J]. Analysis and Testing Technology and Instruments, 1996, (1): 50-51.

岛津IR—408红外分光光谱仪故障二则

Two Special Trouble Examples in Shimadzu IR-408 Infrared Spectrophotometer

  • 摘要: 故障一:开机后扫描起始位置为4000m-1时,记录笔不归零,调“100%满度”控制旅钮也无任何反应,测任何样品记录笔均不动作;按下“AUTOSCAN”键时走纸、扫描系统均正常,转换灵敏度选择开关及用手遮挡样品窗时记录笔均有响应。

     

/

返回文章
返回