HUANG Jin, ZHU Liang, HE Shu-yan. Application of Modified Plane Grating Direct-Reading Spectrometer in Determination of Silver, Boron and Tin in Geochemical Samples[J]. Analysis and Testing Technology and Instruments, 2016, 22(3): 135-139. DOI: 10.16495/j.1006-3757.2016.03.002
Citation: HUANG Jin, ZHU Liang, HE Shu-yan. Application of Modified Plane Grating Direct-Reading Spectrometer in Determination of Silver, Boron and Tin in Geochemical Samples[J]. Analysis and Testing Technology and Instruments, 2016, 22(3): 135-139. DOI: 10.16495/j.1006-3757.2016.03.002

Application of Modified Plane Grating Direct-Reading Spectrometer in Determination of Silver, Boron and Tin in Geochemical Samples

  • Plane grating spectrograph is commonly used in the determination of silver (Ag), boron (B) and Tin (Sn) in geochemical samples. In this work, a CCD-Ⅰplane grating direct-reading spectrometer has been successfully developed to replace the two-meter plane grating spectrograph so as to improve the efficiency and quality of analysis. The detection limits of the updating instrument for Ag, B and Sn were 0.013, 0.71 and 0.43 μg/g, respectively. The relative standard deviations (RSDs n=12) were less than 10%.The analytical accuracy was validated by the national standards and the results are in agreement with the certified values.The proposed method can meet the requirements for the determination of Ag, B and Sn in geochemical samples.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return