XIE Chao, ZHANG Song-hui, CHEN Fei, LI Xue-mei, FAN Xiang-juan. Optimization of Conditions of Inductively Coupled Plasma Mass Spectrometer for Determination of Uranium Elements in Coal Gangue[J]. Analysis and Testing Technology and Instruments, 2017, 23(3): 180-183. DOI: 10.16495/j.1006-3757.2017.03.007
Citation: XIE Chao, ZHANG Song-hui, CHEN Fei, LI Xue-mei, FAN Xiang-juan. Optimization of Conditions of Inductively Coupled Plasma Mass Spectrometer for Determination of Uranium Elements in Coal Gangue[J]. Analysis and Testing Technology and Instruments, 2017, 23(3): 180-183. DOI: 10.16495/j.1006-3757.2017.03.007

Optimization of Conditions of Inductively Coupled Plasma Mass Spectrometer for Determination of Uranium Elements in Coal Gangue

  • A method for the determination of uranium content in coal gangue by inductively coupled plasma mass spectrometry (ICP-MS) was developed. Three sample pretreatment methods were studied and discussed. The instrument conditions were also optimized. The detection limit was 0.05 μg/g, the relative standard deviation (RSD) was less than 4%, and the recovery was between 97% and 103%.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return