Design and Application of an in situ FTIR System for Heterogeneous Catalytic Reaction

  • An in situ FTIR system which can endure hig hreaction temperature for heterogeneous catalytic reaction is desigened and set up. The reaction temperature of the systemform can be from room temperature to 1 150 K,with vacuum from 1.0×105Pa to 1.0×10-4 Pa, and the IR spectral range is 4800cm-1 to 200cm-1. The sample can be easily loaded. The function of the Bruker IFS 120HR FTIR spectrometer is extended and the efficiency of the measurement was also improved.