DONG Xue-zhi, ZHAO Hai-lan, WEI Lin-heng, WU Jing, LIU Xiao-man, AI Tian-zong. Preparation of Ofloxacin Selective Electrode[J]. Analysis and Testing Technology and Instruments, 2001, (2): 78-80.
Citation: DONG Xue-zhi, ZHAO Hai-lan, WEI Lin-heng, WU Jing, LIU Xiao-man, AI Tian-zong. Preparation of Ofloxacin Selective Electrode[J]. Analysis and Testing Technology and Instruments, 2001, (2): 78-80.

Preparation of Ofloxacin Selective Electrode

  • The ofloxacin selective electrode prepared by coating the mixture of ofloxacin and sodium tetraphenylboron association complex with PVC membrane on a glass electrode is reported. The electrode gaved Nernst response to ofloxacin over the concentration range of 5.0×10-3~5.0×10-5 mol/L with a slope of 51.8 mV/pC. The electrode showed a high selectivity and fast response. It has been used for the determination of ofloxacin in ofloxacin injection and capsule. The procedures of this method are simple and the results obtained agree with those of high performance liquid chromatography.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return