MA Jun, TANG Yuan-ming, ZHANG Qing-ying, GONG You-jin. Determination of Hydrogen in Mixing Gas Using Solid-PhaseMicroextraction and Gas Chromatography[J]. Analysis and Testing Technology and Instruments, 2007, (3): 182-186.
Citation: MA Jun, TANG Yuan-ming, ZHANG Qing-ying, GONG You-jin. Determination of Hydrogen in Mixing Gas Using Solid-PhaseMicroextraction and Gas Chromatography[J]. Analysis and Testing Technology and Instruments, 2007, (3): 182-186.

Determination of Hydrogen in Mixing Gas Using Solid-PhaseMicroextraction and Gas Chromatography

  • A method for fast analysis of minimum hydrogen from mixture gas was established. This method was performed by using solid phase microextraction with a preparate Pd-5A molecule sieve, coated fiber coupled with gas chromatography. The results showed that the linear range of hydrogen was 0.001 1~0.125 0 mL/L. The limit detection is 9×10-4 mL/L. This method compared to gas chromatography has the advantages of high sensitivity and less interference. The standard deviations of two methods are 8.7%.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return