JIA Lin, ZHANG Gao, HU Ling, LIU Hong-ni, YANG Cai-ning, WANG Jing-rong. Determination of the Components in the Globule Gun-propellantUsing Accelerated Solvent Extraction(ASE) Followed by HPLC[J]. Analysis and Testing Technology and Instruments, 2008, 14(4): 222-225.
Citation: JIA Lin, ZHANG Gao, HU Ling, LIU Hong-ni, YANG Cai-ning, WANG Jing-rong. Determination of the Components in the Globule Gun-propellantUsing Accelerated Solvent Extraction(ASE) Followed by HPLC[J]. Analysis and Testing Technology and Instruments, 2008, 14(4): 222-225.

Determination of the Components in the Globule Gun-propellantUsing Accelerated Solvent Extraction(ASE) Followed by HPLC

  • Objective: to establish the method for determining the components in globule gun-propellant samples. Method: the samples were extracted by ASE200 and determinated by HPLC with UV detection. Parameters such as solvent, temperature, pressure and time were investigated thoroughly. The ASE conditions were as follows: dichloromethane, 100 ℃ and two 10 min cycles were chosen as the extraction solvent, the temperature and time, respectively. The HPLC separation was performed on a YWG C18 column (150×4.6 mm,10 μm) with methanol and water as the mobile phase at a flow rate of 1.0 mL/min. The detection wavelength was 210 nm. Results: the average recovery of DIANP, NG and C2 were 99.6%, 100.3% and 99.4%, with a RSDof 0.7%,0.8% and 0.9% (n=5). The limits of detection were 2.1 mg/L、1.5 mg/L、0.2 mg/L , respectively. The linear range were 0.02~0.98 g/L, 0.03~1.38 g/L, 0.002~0.124 g/L, respectively. The method was applied to one samples for 5 times.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return