YANG Zhi-qiang, HAO Zhi-jun, WANG Wei, ZENG Ji-jun, ZHANG Wei, LU Jian. Development of New High Accuracy PVT Measurement System[J]. Analysis and Testing Technology and Instruments, 2009, 15(4): 241-244.
Citation: YANG Zhi-qiang, HAO Zhi-jun, WANG Wei, ZENG Ji-jun, ZHANG Wei, LU Jian. Development of New High Accuracy PVT Measurement System[J]. Analysis and Testing Technology and Instruments, 2009, 15(4): 241-244.

Development of New High Accuracy PVT Measurement System

  • A new experimental system of PVT measurement automatically controlled by a computer has been developed. The overall pressure and temperature uncertainty is about ±1.2 KPa, and ±5 mK, respectively. The vapor pressures of HFC-245fa were measured in the temperature range from 294 to 351 K with the new experimental system, and the experimental results indicated that the average deviation between experimental data and the equation provided in the literature was below 0.03%, and the standard deviation was below 0.484%. Which approved that the new experimental system is reliable.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return