RUAN Xin, LI Xiu-yong, ZHANG Fei-fei, LI Pei-nuan, YIN Da-lu, CAO Peng. Determination of Trace Antimony in Water by HydrideGeneration Atomic FluorescenceSpectrometry with Optimization by Orthogonal Experiments[J]. Analysis and Testing Technology and Instruments, 2009, 15(4): 245-250.
Citation: RUAN Xin, LI Xiu-yong, ZHANG Fei-fei, LI Pei-nuan, YIN Da-lu, CAO Peng. Determination of Trace Antimony in Water by HydrideGeneration Atomic FluorescenceSpectrometry with Optimization by Orthogonal Experiments[J]. Analysis and Testing Technology and Instruments, 2009, 15(4): 245-250.

Determination of Trace Antimony in Water by HydrideGeneration Atomic FluorescenceSpectrometry with Optimization by Orthogonal Experiments

  • Using HCl as the medium, the determination of antimony in production water by sequence injection- hydride generation atomic fluorescence spectrometry has been reported. Operating conditions of the instrument were evaluated by orthogonal experiments. Effects of acidity, KBH4 concentration, and thiourea-ascorbic acid content, etc. on measurement results were studied. The effect of coexistent elements was also investigated. Under optimum conditions, the detection limit was 0.011 ng/mL with a linear range of 0~8 ng/mL. The relative standard deviation was 2.78%~6.20%, the recovery rate was 91.0%~99.1%. The method is simple, sensitive and accurate. No significant difference appears between this method and the microwave digestion method.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return